Paramedic Papers

Home Cardiac arrest Pad application is the AED time bottleneck

Cardiac arrest

Pad application is the AED time bottleneck

Identifying the Operational Bottleneck to Shock Delivery After Automated External Defibrillator Activation
Resuscitation · 2026-10 · Added 22 October 2026

Timestamps from 8,115 remotely monitored Japanese AEDs showed pad application consumed a median 81% of the power-on-to-first-analysis interval, with completion inside about 50 seconds marking shocks delivered within 60 seconds of power-on.

Study type
Cohort
Design
Retrospective analysis of remotely monitored AED timestamps
Participants
8,115 AED activation events (900 initially shock-advised with recorded shock)
Main finding
Median power-on-to-pad-application was 42 seconds, accounting for 81% of the power-on-to-first-analysis interval; a threshold of ≥51 seconds identified shock delivery beyond 60 seconds
Caveat
Device log data only, with no linkage to patient outcomes or rescuer type
Bottom line
Hypothesis-generating

Why it matters in Australia (High relevance). Public access defibrillation performance is the same problem everywhere — the device workflow, not the health system, is the bottleneck.

Learning goal

Strengthen my ability to minimise delays between AED power-on and pad application during cardiac arrest resuscitation.

Open in dashboard Read the paper (DOI) PubMed