Home Airway & ventilation Ventilation feedback devices inaccurate during CPR
Airway & ventilation
Ventilation feedback devices inaccurate during CPR
Accuracy of ventilation feedback devices compared with a reference standard during asynchronous ventilation with continuous chest compressions in a simulated cardiac arrest: a prospective non-randomized cross-over study
Resuscitation Plus · Added 8 September 2026
Simulation cross-over study of two commercial ventilation feedback devices used during continuous compressions with asynchronous ventilations — both were clinically inaccurate, one displaying 111 breaths/min when the true rate was 12 and underestimating tidal volume by up to 440 mL in reversed airflow.
- Study type
- Simulation
- Design
- Prospective, simulation-based, non-randomised cross-over study
- Participants
- 20 emergency department nurse practitioners ventilating an intubated manikin (254 and 222 ventilations analysed)
- Main finding
- EOLifeX® overestimated ventilation rate (111/min displayed vs 12/min true; bias +99 breaths/min) and underestimated tidal volume (104 mL vs 544 mL; bias -440 mL); AccuVent® displayed accurate rate but underestimated tidal volume (360 mL vs 532 mL; bias -172 mL)
- Caveat
- Manikin simulation with reversed airflow conditions; only two devices tested
- Bottom line
- Context only
Why it matters in Australia (Medium relevance). transferability — cautionary tale for relying on feedback displays during asynchronous CPR
Learning goal
Weigh the reliability of ventilation feedback devices when guiding ventilations during continuous chest compressions.
More
- Topic: Airway & ventilation
- Daily scan: 8 September 2026
Related papers on Airway & ventilation
-
VAPOX preoxygenation in ED intubation
Australian retrospective cohort of 116 patients intubated with ventilator-assisted preoxygenation (VAPOX) in two EDs — 78% achieved first-pass success without hypoxaemia, but half had at least one complication and 36% became hypotensive after intubation
-
Prehospital cricothyroidotomy: rare but often successful
Nationwide Danish cohort of all 71 prehospital emergency cricothyroidotomies over 2016–2023 — overall procedural success 87.3% (83.3% in OHCA), aspiration and oropharyngeal bleeding the commonest indications, but the procedure was rare (0.12% of 39,885 OHCAs) and OHCA outcomes were poor (ROSC 25%, 30-day survival 4.5%, 1-year survival 2.3%).
-
Paralysing post-ROSC patients prehospital
UK cohort of advanced paramedics in critical care giving neuromuscular blockade to post-ROSC patients poorly tolerating the tube — 99 patients over a year, all PGD indications met, with no adverse physiological trends in sats, end-tidal CO2, MAP or heart rate.
-
ED intubation success high with videolaryngoscopy
Retrospective study of 1,250 ED intubations showed 93.8% first-pass success with videolaryngoscopy as primary technique, with hypotension and desaturation as most common adverse events.