Reproducible measurement of ventilation rate during cardiopulmonary resuscitation using capnography in out-of-hospital cardiac arrest: a cohort study
Resuscitation Plus · 2026-11 · Added 24 November 2026

A Belgian physician-staffed mobile emergency unit cohort tested a capnography-based method for measuring ventilation rate during CPR with an advanced airway — detector agreement was moderate to good (ICC 0.72–0.82) but ventilation rate showed no association with ROSC (OR 1.09 per 2 breaths/min, 95% CI 0.91–1.30).

Study type
Cohort
Design
Cohort study with capnographic breath detection, physician-staffed mobile emergency unit
Participants
83 CPR segments from 75 adult OHCA resuscitation attempts
Main finding
Moderate-to-good agreement with independent detectors (ICC 0.72–0.82, limits of agreement ±6.8 breaths/min) but no association between ventilation rate and ROSC (OR 1.09 per 2 breaths/min, 95% CI 0.91–1.30)
Caveat
The ROSC analysis was exploratory and powered only to detect an odds ratio of at least 1.27 per 2 breaths/min, so clinically relevant effects cannot be excluded
Bottom line
Hypothesis-generating

Why it matters in Australia (Medium relevance). Capnography is on Australian ambulances and 10 breaths/min is a universal CPR quality target, but this is a measurement-methods paper with an exploratory outcome analysis from a physician-staffed unit.

Learning goal

Weigh the value of capnography-based ventilation rate monitoring during cardiac arrest resuscitation in my practice.