Device damage and malfunction during dual defibrillation or dual electrical cardioversion: a scoping review
Resuscitation · Added 27 July 2026
To map and characterize device damage and malfunction associated with dual defibrillator use in defibrillation and cardioversion, which is increasingly employed in both elective and emergency settings, and to identify knowledge gaps relevant to practice and future research.MethodsWe conducted a scop
- Study type
- Systematic review
- Design
- Scoping review including published studies and regulatory adverse event databases
- Participants
- 29 records (2016–2026): 5 published sources and 24 regulatory adverse event reports
- Main finding
- Eight lasting device damage events and four temporary malfunctions identified across published sources; among 24 regulatory reports, 11 described device damage and 16 described malfunctions. Damage was often not recognised during the index event and found on later testing or self-checks.
- Caveat
- Scoping review relying on heterogeneous published sources and voluntary adverse event reports; no denominator, so incidence cannot be estimated
- Bottom line
- Hypothesis-generating
Learning goal
Recognise the risk of occult defibrillator damage when using dual defibrillation for refractory arrhythmias.
More
- Topic: Cardiac arrest
- Daily scan: 27 July 2026
Related papers on Cardiac arrest
-
EEG-guided early sedation withdrawal post-arrest feasible
Early withdrawal of sedation is feasible and safe in patients with an early favourable EEG following cardiac arrest.
-
AI EEG prognostication after cardiac arrest
Accurate prediction of neurological outcome after cardiac arrest is essential for guiding intensive care decisions.
-
Public AED cost per shock
Analysis of public AED deployment cost-effectiveness found very low usage rates relative to investment — the study examines strategies to improve cost per delivered shock.
-
Rescue breathing better in paediatric drowning arrest
In pediatric drowning OHCA, CO-CPR was associated with worse survival and neurological outcomes than RB-CPR.